8L4450A 64-Bit Digital I/O with Memory and Counter

CONF:COUN:FREQ 1e-3, (@1301)

SENS:COUN:INIT (@1301)

SENS:COUN:FREQ? (@1301)

SENS:COUN:PER? (@1301)

SENS:COUN:PWID? (@1301)

SENS:COUN:DCYC? (@1301)

The CONFigure:COUNter:FREQuency command parameter sets the internal gate time (to 1e- 3 or 1 ms in the above example). You can also set the gate time using the SENSe:COUNter:GATE:TIME command.

Clock

The general- purpose clock output is derived from the internal time base. The output clock is divided down from the time base clock such that:

Clock Output (Hz) = (time base frequency)/(divisor)

The time base frequency is 40 MHz. The divisor can be an integer from 2 to 46 providing a range of 20 MHz to 10 Hz for the clock output. The valid values for the clock output rate are: 20 MHz, 13.33 MHz,

10 MHz, 8 MHz, 6.667 MHz, ... 10Hz. The clock output frequency will round to the nearest achievable frequency.

The commands used to control the clock output are:

SOUR:MOD:CLOC:FREQ {<freq>MINMAXDEF},<slot>

SOUR:MOD:CLOC {OFFON01},<slot>

You can obtain the rounded value of the currently set clock frequency using the following query.

SOUR:MOD:CLOC:FREQ?

You can also set the logic “high” voltage level for external clock output. For example, the following command sets the output clock level to 4.5 V.

SOUR:MOD:CLOC:LEV 4.5, 1

L4450A D-Sub Connectors

The L4450A uses two D- sub 78- pin female connectors. Each connector provides contains one bank of the module. As viewed from the rear panel, the connectors and their banks are shown below.

P1 (Bank 1)

P2 (Bank 2)

218

L4400 User’s Guide

Page 228
Image 228
Agilent Technologies L4400 manual Clock, L4450A D-Sub Connectors, P1 Bank P2 Bank