L4451A and L4452A Calibration Procedures

A

4Connect channel 6 of the L4452A to the DMM as shown in Figure A-3. Set the DMM to measure DC volts.

5Set the first DAC calibration point using the command:

CALibration:BEGin:VOLTage 1, (@<channel>)

For channel 6, the command is executed as:

CALibration:BEGin:VOLTage 1, (@1006)

6Measure the DAC output on the DMM. Subtract the offset voltage measured in Step 2 from the reading. Write the result to the DAC using the command:

CALibration:POINt? <value>

Note the DAC output which is now calibration point 2. Subtract the offset voltage from the new DMM reading and write the value to the DAC using:

CALibration:POINt? <value>

7Connect channel 7 of the L4452A to the DMM as shown in Figure A-3. Repeat Steps 5 and 6 for channel 7.