7L4445A Microwave Switch/Attenuator Driver

SCPI Programming Examples

These programming examples provide you with SCPI command examples to use for driving the microwave switch modules.

The channel addressing scheme used in these examples follow the form 1rcc where r is the remote module number (1 through 8), and cc is the two-digit channel number. For more information about channel numbering, refer to “Channel Numbering” on page 143.

For complete information on SCPI commands, see the Programmer’s

Reference Help file.

Example: Configuring an Agilent N1810UL

The following example illustrate controlling an Agilent N1810UL attached to a Y1150A distribution board. The distribution board is connected to Bank 1 of the first remote module attached to the L4445A. This example uses the bank preset (described on page 153).

ROUTe:RMODule:DRIVe:SOURce OFF,(@1100)

ROUTe:RMODule:BANK:PRESet BANK1,(@1100)

ROUTe:RMODule:DRIVe:SOURce INT,(@1101)

ROUT:CLOSe (@1101)

Example: Configuring a Paired Drive Channel

The following example illustrates the sequence of commands used to configure a paired drive channel. In the example, the operations are directed to channel 1 on remote module 3.

The drive source must be disabled before configuring either the channel pairing or the pulse mode. The channel is then paired and the pulse width set to 15 ms. Power supply recovery time and settling time is then set to 12 ms and 10 ms, respectively. Verify is then enabled. The default behavior for the switches is set to OPEN and TTL drive using an EXTernal power supply. Finally, the channel is closed.

ROUTe:RMODule:DRIVe:SOURce OFF,(@1300)

ROUTe:CHANnel:DRIVe:PAIRed ON,(@1301)

ROUTe:CHANnel:DRIVe:PULSe 0.015,(@1301)

ROUTe:CHANnel:DRIVe:TIME:SETTle 0.012,(@1301)

ROUTe:CHANnel:DRIVe:TIME:RECovery 0.010,(@1301)

ROUTe:CHANnel:VERify ON,(@1301)

ROUTe:CHANnel:DRIVe:OPEN:DEFault (@1301)

ROUTe:RMODule:BANK:DRIVe:MODE TTL,BANK1,(@1300)

ROUTe:RMODule:DRIVe:SOURce EXT,(@1300)

ROUT:CLOSe (@1301)

188

L4400 User’s Guide

Page 198
Image 198