4L4421A 40-Channel Armature Multiplexer

34921T Terminal Block

This terminal block with screw- type connections is labeled with the model number and the abbreviated module name.

N O TE

All modules that connect to the analog bus are interlock protected.

This means that when an installed module is exposed

 

 

(no terminal block or cable is connected), the analog bus relays are

 

open and disconnected from the analog buses. See page 100 for

 

further information.

 

 

The 34921T is the only terminal block that provides an isothermal block with temperature reference for thermocouple measurements.

The temperature sensor is located on the bottom side of the PC board as shown below. Also shown are two holes that you can use for connecting an external temperature reference to the terminal block.

Temperature

Sensor

External

Reference

34921T (viewed from bottom side)

C A U T I O N

When wiring the terminal block via cables to the mainframe, make

sure the cables are connected to the correct connector. The cables

 

 

provide communication and power to the temperature sensor on

 

the 34921T terminal block. If cabling is not correct, an error may

 

occur indicating that the L4421A module is not fully operational.

 

 

108

L4400 User’s Guide

Page 118
Image 118
Agilent Technologies L4400 34921T Terminal Block, This means that when an installed module is exposed, Further information