3Operating and Programming

If you define a scan list with the sequential mode enabled and later disable the mode, the scan list will not be reordered; however, the scan list will be treated as a non- sequential list thereafter.

If you have defined a scan list with the sequential mode disabled (OFF) and later enable the mode, the channels will be reordered.

Non- sequential scan lists are not stored as part of the instrument state by the *SAV command; in this case, the ordered mode will be enabled and the scan list will be empty when the instrument state is restored (*RCL command).

The scan order setting is stored in volatile memory and the ordered mode will be enabled when power is turned off or after a Factory Reset (*RST command).

The command used to control the order of the scan list is:

ROUTe:SCAN:ORDered {OFFON}

Monitor Mode

Usage:

L4450A 64- Bit Digital I/O (digital input, counter channels only)

L4452A Multifunction Module (digital input, totalizer channels only)

In the Monitor mode, the instrument takes readings as often as it can on a single channel or during a scan. This feature is useful for troubleshooting your system before a test or for watching an important signal.

Any channel that can be “read” by a device can be monitored. This includes the digital input, totalizer, and the counter channels.

Readings acquired during a Monitor are not stored in memory. However, all readings from a scan in progress at the same time are stored in memory.

The Monitor mode is equivalent to making continuous measurements on a single channel with an infinite scan count. Only one channel can be monitored at a time but you can change the channel being monitored at any time.

A scan in progress always has priority over the Monitor function.

Channels do not have to be part of an active scan list to be monitored; however, the channel must be configured for a measurement in order to be monitored.

The Monitor mode ignores all trigger settings and takes continuous readings on the selected channel using the IMMediate (continuous) source.