5L4433A Dual/Quad 4x8 Reed Matrix

Example: Querying channels for open or close state The following command

returns a 1 (true) or 0 (false) state of channel 204.

ROUTe:CLOSe (@1204)

ROUTe:CLOSe? (@1204) !Returns a 1

ROUTe:OPEN? (@1204) !Returns a 0

Example: Querying the system for module Identify The following command returns the identity of the L4433A:

SYSTem:CTYPe? 1

NOTE

For the L4433A, the query response may include a suffix to indicate

a 1-wire configuration. For example, the response for the L4433A

 

 

will be either "L4433A" (2-wire mode) or "L4433A-1W" (1-wire

 

mode).

 

 

Reading Cycle Count and Resetting the Power-On State

Example: Reading the cycle count for a relay The following command returns

the cycle count on channels 304 and 308.

DIAGnostic:RELay:CYCLes? (@1304,1308)

Example: Resetting the power-on state The following command sets the

L4433A to its power- on state.

SYSTem:CPON 1

Linking Multiple L4433A Instruments

You can link multiple L4433A instruments to form a larger matrix. The following two drawings show two- module connections through rows and columns.

You can connect rows in separate instruments using external wiring. Or, using Bank 2 matrices, you can connect through the analog busses. For a clear idea of how matrices are arranged and their connections to the analog busses, see the simplified schematics later in this chapter. You must use external wiring whenever you connect:

rows in Matrix 1 of separate modules

rows in Matrix 1 to rows in Matrix 2 on the same or separate modules

columns of two matrices on the same or separate modules

116

L4400 User’s Guide

Page 126
Image 126
Agilent Technologies L4400 Linking Multiple L4433A Instruments, Reading Cycle Count and Resetting the Power-On State, Mode