NioWatch operation
c d
BARCO | BARCO |
(1)Reference display
(2)Calibrated display: Luminance automatically matched to reference display
Figure 44: Dual-head equalization
The Equalization option will be available only if:
•the system contains at least one display of the same type as the one being calibrated
•at least one of the displays of the same type has already been calibrated before using a sensor.
To equalize the displays:
1.Be sure the reference display is calibrated to the desired luminance.
2.Set the Equalization option in the Application Settings dialog.
3.Close the Application Settings dialog.
4.Open the Display Settings dialog and calibrate display (2).
The display will be calibrated and additionally the display luminance will be matched to the luminance of the reference display (1).