NioWatch operation

c d

BARCO

BARCO

(1)Reference display

(2)Calibrated display: Luminance automatically matched to reference display

Figure 44: Dual-head equalization

The Equalization option will be available only if:

the system contains at least one display of the same type as the one being calibrated

at least one of the displays of the same type has already been calibrated before using a sensor.

To equalize the displays:

1.Be sure the reference display is calibrated to the desired luminance.

2.Set the Equalization option in the Application Settings dialog.

3.Close the Application Settings dialog.

4.Open the Display Settings dialog and calibrate display (2).

The display will be calibrated and additionally the display luminance will be matched to the luminance of the reference display (1).

System manual E1

83

Page 82
Image 82
Barco E1 manual To equalize the displays, Dual-head equalization