NioWatch operation

Figure 41: DICOM Optimizer dialog

3.Adjust the slider until there is almost no visible difference between the background of the image above the slider and the bitmap inside this image.

4.When done, click button Set point 1 of 20.

5.Repeat this procedure until the button is grayed. This indicates the last point is set.

6.Click OK to finish.

To reset to the latest calibration:

1.Follow step 2 and 3 of the calibration with Barco LCD sensor (see page 77).

2.In the Calibration dialog (see step no. 4. in the description of the calibration with sensor), check the option Use the latest calibration.

80

System manual E1

Page 79
Image 79
Barco E1 manual To reset to the latest calibration, Dicom Optimizer dialog