NioWatch operation

1.Select the desired test pattern by clicking on the corresponding thumbnail or selecting a pattern by name from the Test pattern drop-down box.

2.Click Show to display the pattern.

If you have selected Custom Image, you can select a bitmap image (e.g., saved on the hard disk) as test pattern.

3.To hide the test pattern again, click inside the pattern.

Application settings

Calibration tab

Figure 43: Calibration tab

Here you can check the Equalization option, allowing you to match all the displays from a multi-head system to the luminance of the first calibrated display of the system.

Suppose you work with a system containing 2 displays of the same type, and you wish display (2) to have the same output luminance as display (1). Therefore first calibrate display (1). Next, calibrate display (2) while the Equalization option is checked.

The first calibrated display in the system is the reference display.

82

System manual E1

Page 81
Image 81
Barco E1 manual Application settings, Calibration tab