Sweep height? 370

Configuring the System

Remember if your protocol requires sweeping, that must first be specified in the plate definition used in the protocol.

11.Select YES to specify a sweep on all aspiration cycles. The display will read:

Sweep height? 370

SHOW

Note: The default value is displayed. Select SHOW to move the wash head to the default sweep position so that the height can be checked. Then, determine the correct value as outlined on page 30. Repeat the procedure as needed.

12.Enter the desired value using the keypad and press ENTER. The display will read:

Sweep stroke? 100

Note: There isn’t a SHOW function for the sweep stroke. Determine optimum sweep stroke setting by starting with a small number and increase it if the residual aspirate volume is not acceptable. If you run a protocol and the wash pins strike the sides or bottom of the well, reduce the sweep stroke and review your plate setup parameters.

13.Enter the desired value using the keypad and press ENTER. The display will read:

Bottom wash?

NO

YES

NO

14.Select YES if a bottom wash is to be included. The display will read:

Wash height? 368

SHOW

Note: The default value for bottom wash height is displayed. Select SHOW to move the wash head to the default bottom wash position so that the height can be checked. Then, determine the correct value as outlined on page 30. Repeat the procedure as needed.

Opsys MW™ User’s Guide