Dynex 91000051 manual Operational Problems, Symptom, Probable Cause, Resolution

Models: 91000051

1 95
Download 95 pages 55.29 Kb
Page 86
Image 86
Manual background

Operational Problems

 

Symptom

 

Probable Cause

 

Resolution

 

 

 

 

 

 

 

 

 

Liquid is not

 

The wash

 

Check the system

 

 

 

 

 

 

 

 

 

 

dispensed from the

 

protocol is

 

configuration and wash

 

 

dispense pins

 

incorrectly

 

protocol for the number

 

 

(continued).

 

defined.

 

fluids and fluid volume.

 

 

 

 

The dispense

 

Verify that the dispense

 

 

 

 

 

 

 

 

 

tubing or the

 

tubing and the dispense

 

 

 

 

dispense pump

 

power supply are

 

 

 

 

power supply is

 

connected (see page 20).

 

 

 

 

not connected.

 

 

 

 

 

 

The dispense

 

Remove the tubing from

 

 

 

 

 

 

 

 

 

tubing is sealed

 

the dispense valve and

 

 

 

 

shut at the

 

massage the tubing to

 

 

 

 

dispense valve.

 

release it. Replace the

 

 

 

 

 

 

tubing into the dispense

 

 

 

 

 

 

valve at a different

 

 

 

 

 

 

location.

 

 

 

 

 

 

Purge the tubing more

 

 

 

 

 

 

frequently.

 

 

 

 

 

 

Replace the tubing.

 

 

 

 

The dispense

 

Service is required.

 

 

 

 

 

 

 

 

 

valve does not

 

 

 

 

 

 

open.

 

 

 

 

 

 

The wash head

 

Remove the wash head

 

 

 

 

 

 

 

 

 

manifold tubing is

 

assembly and clean it

 

 

 

 

blocked.

 

using the manifold head

 

 

 

 

 

 

cleaning wire. Then run a

 

 

 

 

 

 

purge. Service is required

 

 

 

 

 

 

if problem persists.

 

 

 

 

There is a

 

Check the seals and

 

 

 

 

pressure leak.

 

tubing for leaks.

 

 

 

 

 

 

Check for air leaks.

 

 

 

 

 

 

Reseat the tubing in the

 

 

 

 

 

 

dispense valve.

 

 

 

 

The dispense

 

Service is required.

 

 

 

 

 

 

 

 

 

pump is faulty.

 

 

 

 

Liquid is dispensed

 

The dispense

 

Insert the dispense tubing

 

 

continuously from

 

tubing is not

 

into the dispense valve

 

 

the dispense pins

 

inserted in the

 

(see page 20).

 

 

 

 

dispense valve.

 

 

 

 

 

 

 

 

 

 

78

Opsys MW™ User’s Guide

Page 86
Image 86
Dynex 91000051 manual Operational Problems, Symptom, Probable Cause, Resolution