Dynex 91000051 manual Wash Protocols Wash Protocol Operations, Purge, Move, Aspirate, Description

Models: 91000051

1 95
Download 95 pages 55.29 Kb
Page 47
Image 47
Wash Protocol Operations

Wash Protocol Operations

Chapter 4 Wash Protocols

Wash Protocol Operations

Awash protocol consists of a series of Purge, Move, Soak, Aspirate, Dispense and Fill operations.

Purge, Move and Soak can be carried out in any sequence. Aspirate, Dispense and Fill can only be carried out within a Move operation, and there cannot be a Move within a Move.

Each of these operations is described below:

 

Operation

 

Description

 

 

 

 

Dispenses fluid from the dispense wash pins while

 

 

 

 

 

 

Purge

 

 

 

 

 

the wash head is positioned over the purge tray

 

 

 

 

Note: Purge is usually

 

 

 

 

carried out at the beginning

 

 

 

 

of a wash protocol to rinse

 

 

 

 

the dispense wash pins and

 

 

 

 

remove air bubbles.

 

 

Move

 

Performs Aspirate, Dispense, Fill and/or Soak

 

 

 

 

 

 

 

operations on specified strips of the Microplate

 

 

Aspirate

 

Removes the contents of a well by positioning the

 

 

 

 

wash pins at the aspiration height in the well and

 

 

 

 

aspirating the liquid from the wells

 

 

 

 

A sweep may also be performed during Aspirate.

 

 

 

 

Note: The aspirate height

 

 

 

 

and sweep parameters for a

 

 

 

 

particular plate type are

 

 

 

 

specified during system

 

 

 

 

configuration (30).

 

 

 

 

If a sweep is specified during

 

 

 

 

configuration, the wash pins

 

 

 

 

are moved back and forth

 

 

 

 

during the aspiration cycle.

 

Opsys MW™ User’s Guide

39

Page 47
Image 47
Dynex 91000051 manual Wash Protocols Wash Protocol Operations, Purge, Move, Performs Aspirate , Dispense , Fill and/or Soak