Dynex 91000051 manual Pump Calibration

Models: 91000051

1 95
Download 95 pages 55.29 Kb
Page 83
Image 83
8.If the dispensed volume is too high, decrease the pump speed using the SETUP/CAL function (refer to page 36).

Pump Calibration

8.If the dispensed volume is too high, decrease the pump speed using the SETUP/CAL function (refer to page 36).

A pump speed change of -10 units roughly corresponds to a well volume change of -10µl.

9.If the dispensed volume is too low, increase the pump speed using the SETUP/CAL function (refer to page 36).

A pump speed change of +10 units roughly corresponds to a well volume change of +10µl.

10.Repeat Steps 3-9 until fluid is dispensed with the accuracy required. (e.g., Error less than or equal to ±5%)

Opsys MW™ User’s Guide

75

Page 83
Image 83
Dynex 91000051 manual Pump Calibration