Preparation

Preparation

Chapter 5 Washing a Plate

Once the parameters for at least one plate type and one wash protocol have been set, the system can be used to wash a plate.

Preparation

The only preparation that is required is to check the fluid levels in the Wash Container and the Waste Container. Service the containers if necessary.

To empty the waste container:

1.Unscrew the cap of the Waste Container. Lay the cap and level sensor assembly on a paper towel.

2.Discard the waste in accordance with local regulations.

3.Rinse the waste container with deionized water. Discard the rinse water.

Note: If desired, the waste container can be disinfected with a 10% (v/v) solution of household bleach in water.

4.Place the level sensor and tube back into the Waste Container and screw the cap on.