Dynex 91000051 manual Configuring the System

Models: 91000051

1 95
Download 95 pages 55.29 Kb
Page 40
Image 40
Note: The default value is displayed. Select SHOW to move the wash head to the default well top position so that the height can be checked. Then, determine the correct value as outlined on page 30. Repeat the procedure as needed.

Configuring the System

7.Enter the dispense height using the keypad and press ENTER. The display will read:

Well top? 160

SHOW

Note: The default value is displayed. Select SHOW to move the wash head to the default well top position so that the height can be checked. Then, determine the correct value as outlined on page 30. Repeat the procedure as needed.

8.Enter the desired value using the keypad and press ENTER. The display will read:

ASP. HEIGHT? 380

SHOW

Note: The default value is displayed. Select SHOW to move the wash head to the default aspiration position so that the height can be checked. Then, determine the correct value as outlined on page 30. Repeat the procedure as needed.

9.Enter the desired value using the keypad and press ENTER. The display will read:

Sweep? NO

YESNO

10. Select YES to specify a sweep. The display will read:

All cycles?

NO

YES

NO

Note: If a sweep is specified, the wash pins are moved up and down and across the bottom of the well during aspiration to better remove all droplets of liquid. If NO is selected, the Wash height? prompt (Step 14) will be displayed.

Note: Use the sweep function with Flat or “C” bottom plates only. Sweeping aspirates fluid more completely from the wells than a regular aspiration. Sweep all cycles to get the most thorough well aspiration when using multiple aspiration cycles, but remember that each sweep adds seconds to the protocol.

32

Opsys MW™ User’s Guide

Page 40
Image 40
Dynex 91000051 manual Configuring the System