Specifying the Pump Parameters

Configuring the System

Specifying the Pump Parameters

The CAL. command allows the calibration values for each pump to be specified:

1. Select SETUP (if needed). The display will read:

Select option

HEAD PLATE

2. Scroll right to display the next selection:

Select option

VERSION CAL.

3. Select the pump pressure (high or low). The display will read:

Pump speed?

190

 

 

4.Enter the desired value using the keypad and press ENTER. The display will read:

Slope?2.1

Note: The Slope and Intercept values should only be changed by DYNEX Technologies Service Personnel.

5. Press ENTER to keep the existing value. The display will read:

Intercept? 24

6.Press ENTER to keep the existing value. The display will again read:

Select option

VERSION CAL.

Note: Refer to page 74 for additional pump calibration instructions.