Cleaning the Containers

Cleaning the Containers

Cleaning the Containers

Procedures for emptying, cleaning and refilling the Wash Container and the Waste Container are described below.

To empty and clean the waste container:

1.Unscrew the cap of the Waste Container. Lay the cap and level sensor assembly on a paper towel.

2.Discard the waste in accordance with local regulations.

3.Rinse the waste container with deionized water. Discard the rinse water.

Note: If desired, the waste container can be disinfected with a 10% (v/v) solution of household bleach in water.

4.Place the level sensor and tube back into the Waste Container and screw the cap on.

To refill the wash container:

1.Separate the dispense line quick-connector in the cap of the Wash Container. Unscrew the cap and lay it and the Pick-Up Filter on a clean paper towel.

2.Fill the Wash Container with the wash fluid that is being used.

Note: If the wash solution is being changed, discard the contents of the Wash Container and thoroughly clean it before filling it with the new wash solution.

3.Place the Pick-Up Filter into the Wash Container and screw the cap on. Reconnect the dispense line quick-connector.

70

Opsys MW™ User’s Guide

Page 78
Image 78