To specify a Purge:

Creating a New Wash Protocol

To specify a Purge:

Note: A purge must be performed outside of a

Move cycle.

1.Display the header information for the selected wash protocol, as outlined on page 41. The display will read:

Select option

MOVE PURGE

2. Select PURGE. The display will read:

Fluid volume?

350

 

 

3. Use the keypad to enter the volume of fluid to be dispensed.

Note: A purge volume ranging from 50 to 9,999 microliters can be entered.

4. Press ENTER. The display will read:

Select option

MOVE PURGE

46

Opsys MW™ User’s Guide

Page 54
Image 54
Dynex 91000051 manual To specify a Purge, Creating a New Wash Protocol