Dynex 91000051 manual Creating a New Wash Protocol, Select plate 2 DYNEX1, Variable strips? YESNO

Models: 91000051

1 95
Download 95 pages 55.29 Kb
Page 51
Image 51
7. Re-enter the password and press ENTER. The display will read:

Creating a New Wash Protocol

7. Re-enter the password and press ENTER. The display will read:

Select plate 2: DYNEX_1

Note: The plate name that is shown is an example. The actual name will be the plate name that is entered for that plate type during system configuration (see page 30).

8.Use the up and down scroll keys to select the plate type (or enter the number using the keypad) to be used and press ENTER. The display will read:

Variable strips?

YESNO

Note: “Variable Strips” is used to run an established protocol on fewer strips than originally specified in the “MOVE” cycle(s) of an established protocol.

If YES is selected, the operator will be prompted at run time to specify the number of strips that are to be washed. Processing will begin on the first strip selected in the “MOVE” cycle and will end on the total number of strips entered at run time (e.g. the protocol states: “Move 3-12 1 cycle(s)” and at run time, 2 strips are entered, only strips 3 and 4 will be processed).

9. If YES is selected, the display will read:

Constant timing?

YESNO

Note: If NO is selected for Variable strips?, the Select option prompt will be displayed (see below).

Note: The Constant Timing prompt only appears if you are using Variable Strips.