Pump Calibration
Pump Calibration
Each Opsys MW™ washer manifold, or head, (e.g., 1 x 8, 1 x 12) has three calibration values stored in memory. These values are calculated and stored during manufacture at DYNEX Technologies. When the Opsys MW configuration is changed from one manifold to another, the corresponding calibration settings will be recalled by correctly setting the head size using the SETUP/HEAD feature (refer to page 28 for Specifying the Wash Head).
This calibration protocol may be used at any time to verify the performance of the Opsys MW washer.
IMPORTANT! It is recommended that this protocol be completed if any
Calibration Protocol:
1.Define a flat bottom plate profile using the SETUP/PLATE function (refer to page 30 for Specifying Plate Parameters).
2.Program an assay to dispense 200µl/well into columns
3.Obtain a clean, dry flat bottom
4.Weigh the plate and record the empty plate weight in grams.
5.Run the assay programmed in Step 2. (This assay will dispense 200µl to all 96 wells in the plate.)
6.Weigh the plate and record the filled plate weight in grams.
7.Calculate the actual dispensed volume and % error using the following formulas:
A.Actual Volume = [(Full weight – Empty weight) x 1000mg/g] / (96 wells x fluid density)
B.%Error = 100 x (Actual Volume – Target Volume)/Target Volume
Example:
Empty plate weight: 48.38g
Full plate weight: 67.42g
Density of water at 20°C = 0.998mg/µl
Actual Volume = [(67.42 – 48.38) x 1000]/(96 x 0.998) = 198.7µl/well
%Error = 100 x (198.7 – 200)/200 =