Dynex 91000051 manual Specifying Wash Cycle Operations, Creating a New Wash Protocol

Models: 91000051

1 95
Download 95 pages 55.29 Kb
Page 53
Image 53
Specifying Wash Cycle Operations

Creating a New Wash Protocol

Specifying Wash Cycle Operations

The final step in creating a wash protocol is to specify individual operations within the wash protocol.

An example of a wash protocol is shown in 45. In this example, the dispense wash pins are purged, and then a Dispense and Aspirate are carried out twice on each of the 12 strips (columns) of a 96-well Microtiter® plate. Finally, after all 12 strips have been washed twice, the contents of the Microtiter plate are allowed to soak (equilibrate) for another 600 seconds.

PURGE

MOVE

(2 Cycles)

Purge the dispense pins.

Specify the starting strip and ending strip for each cycle and the number of cycles to be performed.

DISPENSE

(3 Times)

For each strip, perform dispense three times.

ASPIRATE

For each strip, perform one aspiration cycle.

END LOOP

End the cycle.

SOAK

After the two Move cycles have been completed, allow the contents to equilibrate for 600 seconds.

END

End the wash protocol.

Figure 9. Example of a Wash Protocol

The procedures for specifying the steps in this and other wash protocols are described on the following pages.

Opsys MW™ User’s Guide

45

Page 53
Image 53
Dynex 91000051 manual Specifying Wash Cycle Operations, Creating a New Wash Protocol