To replace the vacuum tubing:

Replacing the Tubing

To replace the vacuum tubing:

Note: This procedure describes replacement of the vacuum tubing connecting the Waste Container to the Foam/Aerosol Trap. The vacuum tubing connecting the Foam/Aerosol Trap to the rear of the instrument should not require changing.

1.Unscrew the Cap of the Waste Container, lay it and the level sensor assembly on a clean paper towel.

2.Discard the contents of the Waste Container in accordance with local regulations. Thoroughly rinse the container with deionized water and let it dry.

3.Disconnect the vacuum tubing from the connector in the Cap and from the connector in the Foam/Aerosol Trap. Discard the old tube.

4.Obtain a new length of vacuum tubing.

5.Connect one end of the vacuum tubing to the connector in the Waste Container Cap and the other end to the connector in the Foam/Aerosol Trap.

6.Place the level sensor assembly back into the Waste Container and screw the Cap on.

72

Opsys MW™ User’s Guide

Page 80
Image 80
Dynex 91000051 manual To replace the vacuum tubing, Replacing the Tubing