Oxymitter 4000

a. Fault 1, Open Thermocouple

The HEATER T/C LED flashes once,

pauses for three seconds, and repeats

(Figure 7-1).

1. Check connector J1. Ensure the con-

nector is properly seated.

TEST

POINTS

INC INC

HIGH LOW

GAS GAS

DEC DEC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS + PROCESS - % O2

TP1 TP2 TP3 TP4

TP5 TP6

J1

REDYELGRNORG

22220033

2.

Using a multimeter, measure TP3+ to

 

TP4-. If the reading is 1.2 VDC ±0.1

 

VDC, the thermocouple is open.

3.

Remove power. Disconnect J1. Meas-

 

ure continuity across the red and yel-

 

low thermocouple leads.

4.

The measurement should read ap-

 

proximately 1 ohm.

5.

If the thermocouple is open, see para-

 

graph 6-7,Heater Strut Replacement.

Figure 7-1. Fault 1, Open Thermocouple

7-4 Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 102
Image 102
Emerson Process Management IB106-340 instruction manual Fault 1, Open Thermocouple