Instruction Manual

Oxymitter 4000

IB-106-340 Rev. 2.4

April, 2001

 

 

 

HEATER T/C

 

DIAGNOSTIC

HEATER

SW2

ALARMS

O2 CELL

ON

 

 

CALIBRATION

 

CALIBRATION RECOMMENDED

HEATER T/C

HEATER

O2 CELL

CALIBRATION

TEST

POINTS

INC INC

HIGH LOW

GAS GAS

DEC DEC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS +

PROCESS -

% O2

TP1 TP2 TP3 TP4

TP5 TP6

J1 RED YEL GRN ORG

1 2 3 4 1 2 3 4

LIGHTING SEQUENCE DURING NORMAL OPERATION

HEATER T/C

HEATER

O2 CELL

CALIBRATION

1

2

3

4

1

2

3

4

 

LIGHTING SEQUENCE DURING WARM-UP

 

22220056

3

Figure 3-3. Startup and Normal Operation

3-3 RECOMMENDED CONFIGURATION

necessary calibrations based on calendar

 

days or weeks since previous calibration.

a. 4-20 mA Signal Upon Critical Alarm

When utilizing the SPS 4000 or the IMPS

 

4000, consider wiring some or all associ-

Rosemount recommends that the factory

ated alarm contacts.

default be utilized. The 4-20 mA signal will

1.

CALIBRATION INITIATE. Contact

go to the 3.8 mA level upon any critical

alarm which will cause the O2 reading to be

 

from the control room to an SPS 4000

unusable. Customer can also select 22 mA

 

or IMPS 4000 (one per probe) provides

as the failure setting if normal operations

 

the ability to manually initiate a calibra-

cause O2 readings to go below the zero %

 

tion at any time from the control room.

O2 (4 mA) level.

 

Note that calibrations can also be initi-

 

 

ated from a HART handheld communi-

If the O2 measurement is being utilized as

 

cator, from Asset Management

part of an automatic control loop, the loop

 

Solutions software, or from the keypad

should be placed into manual upon this fail-

 

on the Oxymitter 4000.

ure event or other appropriate action should

2.

IN CALIBRATION. One contact per

be taken.

 

 

probe provides notification to the con-

b. Calibration

 

trol room that the “calibration recom-

 

 

mended” diagnostic has initiated an

Rosemount recommends utilizing an auto-

 

automatic calibration through the SPS

calibration system, actuated by the “calibra-

 

4000 or IMPS 4000. If the O2 signal is

tion recommended” diagnostic. New O2

 

being utilized in an automatic control

cells may operate for more than a year, but

 

loop, this contact should be utilized to

older cells may require recalibration every

 

place the control loop into manual dur-

few weeks as they near the end of their life.

 

ing calibration.

This strategy ensures that the O2 reading is always accurate, and eliminates many un-

Rosemount Analytical Inc. A Division of Emerson Process Management

Startup 3-5

Page 61
Image 61
Emerson Process Management IB106-340 Recommended Configuration, Necessary calibrations based on calendar, Calibration