Oxymitter 4000

e. Fault 5, Open Heater

The HEATER LED flashes once, pauses for

three seconds, and repeats (Figure 7-5).

1. Remove power. Remove the electronic

assembly per paragraph 6-5b,Elec-

tronic Assembly Replacement.

TEST

POINTS

INC INC

HIGH LOW

GAS GAS

DEC DEC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS + PROCESS - % O2

TP1 TP2 TP3 TP4

TP5 TP6

J1 RED YEL GRN ORG

22220036

2.

Using a multimeter, measure across

 

the heater connector J8.

3.

The measurement should be approxi-

 

mately 72 ohms. If the heater is open,

 

see paragraph 6-7,Heater Strut

 

Replacement.

Figure 7-5. Fault 5, Open Heater

7-8 Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 106
Image 106
Emerson Process Management IB106-340 instruction manual Fault 5, Open Heater