b. Fault 2, Shorted Thermocouple

The HEATER T/C LED flashes twice,

pauses for three seconds, and repeats

(Figure 5-2).

1. Using a multimeter, measure across

TP3+ and TP4-.

TEST

POINTS

INC INC

HIGH LOW

GAS GAS

DEC DEC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS + PROCESS - % O2

TP1 TP2 TP3 TP4

TP5 TP6

J1 RED YEL GRN ORG

22220034

2.

If the reading is 0 ±0.5 mV, then a

 

shorted thermocouple is likely.

3.

Remove power and disconnect J1.

4.

Measure from TP3+ to TP4-. The

 

reading should be approximately 20

 

Kohms.

5.

If so, the short is not on the PC board.

 

See paragraph 6-7,Heater Strut Re-

 

placement.

Figure 7-2. Fault 2, Shorted Thermocouple