n. Fault 14, Invalid Constant

The CALIBRATION LED flashes twice,

pauses for three seconds, and repeats

(Figure 7-14).

1. After a calibration has been performed,

the electronics calculates a cell con-

stant value.

TEST

POINTS

INC INC

HIGH LOW

GAS GAS

DEC DEC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS + PROCESS - % O2

TP1 TP2 TP3 TP4

TP5 TP6

J1

REDYELGRNORG

22220045

2.

If the cell constant value is outside of

 

the range, -4 mV to 10 mV, the alarm

 

will activate. See paragraph 6-2,Cali-

 

bration, and verify the last calibration

 

was performed correctly.

3.

Power down the Oxymitter 4000 and

 

remove it from the stack.

4.

Replace the cell per paragraph 6-8,

 

Cell Replacement.