Instruction Manual

IB-106-340 Rev. 2.4 April, 2001

Oxymitter 4000

QUICK REFERENCE GUIDE

OXYMITTER 4000 OXYGEN TRANSMITTER

Performing a Manual Calibration

1.Place the control loop in manual.

2.Press the CAL key. The CAL LED will light solid.

3.Apply the first calibration gas.

4.Press the CAL key. When the unit has taken the readings using the first calibration gas, the CAL LED will flash continuously.

5.Remove the first calibration gas and apply the second calibration gas.

6.Push the CAL key. The CAL LED will light solid. When the unit has taken the readings using the second calibration gas, the CAL LED will flash a two-pattern flash or a three-pattern flash. A two-pattern flash equals a valid calibration, three-pattern flash equals an invalid cali- bration.

7.Remove the second calibration gas and cap off the calibration gas port.

8.Press the CAL key. The CAL LED will be lit solid as the unit purges. When the purge is complete, the CAL LED will turn off.

9.If the calibration was valid, the DIAGNOSTIC ALARMS LEDs indicate normal operation. If the new calibration values are not within the parameters, the DIAGNOSTIC ALARMS LEDs will indicate an alarm.

10.Place the control loop in automatic.

P-14

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 24
Image 24
Emerson Process Management IB106-340 instruction manual Quick Reference Guide Oxymitter 4000 Oxygen Transmitter