Instruction Manual

IB-106-340 Rev. 2.4 April, 2001

Oxymitter 4000

 

ALARMS

LED

 

FLASHES

STATUS

 

 

1

OPEN

 

 

2

SHORTED

HEATER T/C

3

REVERSED

 

 

 

 

4

A/D COMM

 

 

ERROR

 

 

 

 

 

1

OPEN

 

2

HIGH HIGH

 

TEMP

 

 

HEATER

3

HIGH CASE

TEMP

 

4

LOW TEMP

5

HIGH TEMP

1

OPEN

3

BAD

O2 CELL

 

 

4

EPROM

 

CORRUPT

 

 

 

1

INVALID SLOPE

 

 

 

CALIBRATION

2

INVALID

 

 

CONSTANT

 

 

 

 

 

3

LAST CAL

 

 

FAILED

 

 

 

MANUAL

CALIBRATION

*PLACE CONTROL LOOP IN MANUAL

*IF CAL LED ON GO TO STEP 2

1

PUSH CAL

CAL LED ON

 

2

PUSH CAL

CAL LED FLASH

 

3APPLY TG1 PUSH CAL

4CAL LED ON SOLID WAIT FOR FLASH

5REMOVE TG1 & APPLY TG2

PUSH CAL

CAL LED ON SOLID

6WAIT FOR FLASH

2 FLASH-VALID CAL

3 FLASH-INVALID CAL

7REMOVE TG2

PUSH CAL

CAL LED ON FOR

8PURGE TIME CAL LED OFF

SW2 DIP SWITCH

HART

 

LOCAL

0-10%

 

0-25%

4mA

 

20mA

NOT USED

 

NOT USED

Figure 6-3. Inside Right Cover

Use the following to perform a manual calibration:

(a)Place control loop in manual.

(b)Verify the calibration gas parame- ters are correct (paragraph 6-2a).

(c)If performing a manual calibration with CALIBRATION RECOM- MENDED LED off and the CAL LED off, start at step 1.

(d)If performing a manual calibration with CALIBRATION RECOM- MENDED LED on and the CAL LED on, start at step 2.

1 Push the CAL key. The CALI-

BRATION RECOMMENDED LED will come on and the CAL LED will be on solid. If a multimeter is attached across TP5 and TP6, the reading will display the percentage of oxygen seen by the cell.

29770005

2Push the CAL key. The CALI-

BRATION RECOMMENDED LED will turn off and the CAL LED will flash continuously.

The Oxymitter 4000 can be configured so that the 4-

20 mA signal will hold the last value. The default condition is for the output to track. the flashing LED indicates that the Oxymitter 4000 is ready to accept the first calibration gas.

3Apply the first calibration gas. (Electronics will abort the cali- bration if step 4 is not done within 30 minutes).

4Push the CAL key; the CAL LED will be on solid. A timer is activated to allow the calibra- tion gas adequate time to flow (default time of five minutes). When the timer times out, the Oxymitter 4000 has taken the

6-4 Maintenance and Service

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 78
Image 78
Emerson Process Management IB106-340 instruction manual Alarms, Manual Calibration