Instruction Manual
Oxymitter 4000
NOTE
To determine if the diffusion element needs to be replaced, refer to para- graph
c.If equipped with the optional ceramic diffu- sion assembly, remove and discard the setscrews and remove the vee deflector (Figure
d.Loosen the four socket head cap screws from the cell and flange assembly and re- move the assembly and the corrugated seal. The cell flange has a notch that may be used to gently pry the flange away from the probe. Note that the contact pad inside of the probe will sometimes fuse to the oxy- gen sensing cell. If the cell is fused to the contact pad, push the cell assembly back into the probe (against spring pressure) and quickly twist the cell assembly. The cell and contact pad should separate. If the contact pad stays fused to the cell, a new contact/thermocouple assembly must be in- stalled. Disconnect the cell and the thermo- couple wires at the probe electronics and withdraw the cell with the wires still attached.
e.Remove entire electronics per paragraph
f.If the contact assembly is damaged, replace the strut or the contact pad. Instructions for replacing the contact pad are in the cell re- placement kit.
g.Remove and discard the corrugated seal. Clean the mating faces of the probe tube and retainer. Remove burrs and raised surfaces with a block of wood and crocus cloth. Clean the threads on the retainer and hub.
h.Rub a small amount of
i.Assemble the cell and flange assembly, cor- rugated seal, and probe tube. Make sure the calibration tube lines up with the cali- bration gas passage in each component. Apply a small amount of
j.Install the entire electronics per paragraph
k.Apply
l.On systems equipped with an abrasive shield, install the dust seal gaskets, with joints 180° apart.
m.Reinstall the probe and gasket on the stack flange.
n.Follow the instructions in paragraph
o.Turn on power and monitor thermocouple output. It should stabilize at 29.3+0.2 mV. Set reference air flow at 2 scfh (56.6 l/hr).
After the Oxymitter 4000 stabilizes, calibrate the unit. If new components have been in- stalled, repeat calibration after 24 hours of operation.
Rosemount Analytical Inc. A Division of Emerson Process Management |