SECTION 3 STARTUP

Install all protective equipment covers and safety ground leads before equipment startup. Failure to install covers and ground leads could result in serious injury or death.

3-1 GENERAL

a.Verify Mechanical Installation

Ensure the Oxymitter 4000 is installed cor- rectly (Section 2, INSTALLATION).

b.Verify Terminal Block Wiring

1.Remove screw (32, Figure 6-1),gasket (33), and cover lock (34) that secure the terminal block cover. Remove the cover to expose the terminal block (Figure 3-1).

2.Check the terminal block wiring. Be sure the power, 4-20 mA signal, and logic outputs are properly connected and secure.

3. Install the housing cover on the termi-

3

 

nal block and secure with cover lock

 

(34, Figure 6-1),gasket (33), and

 

screw (32).

 

4.For an Oxymitter 4000 with an inte- grally mounted SPS 4000, remove screws (26, Figure 6-11)and terminal cover (27). Check that the power and signal terminations are properly con- nected to terminal strip (25) and secure according to instructions in Section 2,

INSTALLATION.

5.Install terminal cover (27) and secure with screws (26).

+ 4-20 -

-

+

 

 

 

AC

AC

 

 

N

L1

4-20 mA

 

SIGNAL

TERMINAL

 

BLOCK

LOGIC I/O

 

Rosemount Analytical Inc.

SMART FAMILY

Orrville, OH 44667-0901

800-433-6076

HART

OXYMITTER 4000

SERIAL NO.

TAG NO.

VOLTS: 85-264 VAC

48-62 Hz WATTS:

OUTPUT: 4-20 mA

LINE FUSE:

OXYMITTER 4000

ELECTRONICS

HOUSING

500 VA

5 Amps

HEATER T/C

DIAGNOSTIC HEATER SW2

ALARMS 02 CELL

CALIBRATION

CALIBRATION RECOMMENDED

 

 

02 CELL mV +

TP1

J1

 

 

 

TEST

02 CELL mv -

TP2

 

POINTS

HEATER T/C +

TP3

 

 

 

HEATER T/C -

TP4

RED YEL GRN ORG

 

 

 

 

INC

INC

CAL

 

 

 

 

 

 

HIGH

LOW

 

 

 

GAS

GAS

 

 

 

DEC

DEC

TEST GAS +

TP5

 

PROCESS -

TP6

 

 

 

 

 

 

% 02

 

 

GROUND LUGS

26170036

Figure 3-1. Integral Electronics

Rosemount Analytical Inc. A Division of Emerson Process Management

Startup 3-1

Page 57
Image 57
Emerson Process Management IB106-340 Startup, General, Verify Mechanical Installation, Verify Terminal Block Wiring