Instruction Manual
Oxymitter 4000
The logic contact is | 2. Additional IMPS 4000 Alarm Contacts. | |||
340 ohm series resistance. An interposing | (a) | One contact per IMPS 4000 for | ||
relay will be required if this contact is to be | ||||
utilized to annunciate a higher voltage de- |
| “low calibration gas flowing”. | ||
vice, such as a light or horn, and may also | (b) | One contact per IMPS 4000 for | ||
be required for certain DCS input cards. A | ||||
| “high calibration gas flowing”. | |||
Potter & Brumfield |
| |||
|
| |||
mA DC or an equal interposing relay will be |
| NOTE | ||
mounted where the contact wires terminate |
| |||
The | ||||
in the control/relay room. | ||||
|
| to respond normally during any cali- | ||
b. If autocalibration systems are utilized, the | bration, or can be configured to hold | |||
bidirectional logic contact is utilized as a | the last O2 value upon the initiation of | |||
“handshake” signal between the autocali- | calibration. Factory default is for the | |||
bration system (SPS 4000 or IMPS 4000) | ||||
and is unavailable for alarming purposes. | throughout calibration. Holding the | |||
The following additional contacts are pro- | last O2 value may be useful if several | |||
vided through the autocalibration systems: | probes are being averaged for the | |||
|
| purpose of automatic control. Unless | ||
1. SPS 4000 and IMPS 4000, | several probes are being averaged, | |||
|
| always place any control loops using | ||
(a) | One contact closure per probe | the O2 signal into manual prior to | ||
| from the control room to the SPS | calibrating. |
| |
| 4000 or IMPS 4000 for “calibration |
| ||
| initiate”. | |||
|
| INDICATIONS | ||
(b) | One contact output per probe from | Faults in the Oxymitter 4000 are indicated using | ||
| the SPS 4000 or IMPS 4000 to the | |||
| the four diagnostic, or unit, alarms. The pattern | |||
| control room for “in calibration” | |||
| of repeating blinks will define the problem. A | |||
| notification. | |||
| condensed table of the errors and the corre- | |||
|
| |||
(c) | Once contact output per probe | sponding blink codes can be found on the inside | ||
right cover of the electronics housing. Table |
from the SPS 4000 or IMPS 4000 | also identifies the blink code and fault status of | |
to the control room for “calibration | ||
each LED as well as the output of the | ||
failed” notification. (Includes output | ||
signal line and a fault number that corresponds | ||
from pressure switch indicating “cal | ||
to the troubleshooting instructions provided in | ||
gas bottles empty”). | ||
this section. | ||
|
Rosemount Analytical Inc. A Division of Emerson Process Management |