Instruction Manual

IB-106-340 Rev. 2.4 April, 2001

Oxymitter 4000

The logic contact is self-powered, +5 VDC,

2. Additional IMPS 4000 Alarm Contacts.

340 ohm series resistance. An interposing

(a)

One contact per IMPS 4000 for

relay will be required if this contact is to be

utilized to annunciate a higher voltage de-

 

“low calibration gas flowing”.

vice, such as a light or horn, and may also

(b)

One contact per IMPS 4000 for

be required for certain DCS input cards. A

 

“high calibration gas flowing”.

Potter & Brumfield R10S-E1Y1-J1.0K 3.2

 

 

 

mA DC or an equal interposing relay will be

 

NOTE

mounted where the contact wires terminate

 

The 4-20 mA signal can be configured

in the control/relay room.

 

 

to respond normally during any cali-

b. If autocalibration systems are utilized, the

bration, or can be configured to hold

bidirectional logic contact is utilized as a

the last O2 value upon the initiation of

“handshake” signal between the autocali-

calibration. Factory default is for the

bration system (SPS 4000 or IMPS 4000)

4-20 mA signal to operate normally

and is unavailable for alarming purposes.

throughout calibration. Holding the

The following additional contacts are pro-

last O2 value may be useful if several

vided through the autocalibration systems:

probes are being averaged for the

 

 

purpose of automatic control. Unless

1. SPS 4000 and IMPS 4000, 1-4 probes.

several probes are being averaged,

 

 

always place any control loops using

(a)

One contact closure per probe

the O2 signal into manual prior to

 

from the control room to the SPS

calibrating.

 

 

4000 or IMPS 4000 for “calibration

7-4 IDENTIFYING AND CORRECTING ALARM

 

initiate”.

 

 

INDICATIONS

(b)

One contact output per probe from

Faults in the Oxymitter 4000 are indicated using

 

the SPS 4000 or IMPS 4000 to the

 

the four diagnostic, or unit, alarms. The pattern

 

control room for “in calibration”

 

of repeating blinks will define the problem. A

 

notification.

 

condensed table of the errors and the corre-

 

 

(c)

Once contact output per probe

sponding blink codes can be found on the inside

right cover of the electronics housing. Table 7-1

from the SPS 4000 or IMPS 4000

also identifies the blink code and fault status of

to the control room for “calibration

each LED as well as the output of the 4-20 mA

failed” notification. (Includes output

signal line and a fault number that corresponds

from pressure switch indicating “cal

to the troubleshooting instructions provided in

gas bottles empty”).

this section.

 

7-2 Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 100
Image 100
Emerson Process Management IB106-340 instruction manual Identifying and Correcting Alarm, Indications