Oxymitter 4000

Instruction Manual

IB-106-340 Rev. 2.4 April, 2001

readings using the first cali-

 

8 Push the CAL key; the CAL

bration gas and the CAL LED

 

LED will be on solid as the

will flash continuously. The

 

unit purges. (Default purge

flashing indicates the Oxy-

 

time is three minutes). When

mitter 4000 is ready to take

 

the purge is complete, the

readings using the second

 

CAL LED will turn off and the

calibration gas.

 

Oxymitter 4000 output un-

 

 

locks from its held value and

5 Remove the first calibration

 

begins to read the process O2.

gas and apply the second

 

 

calibration gas. (Electronics

 

If the calibration was valid, the DI-

will abort the calibration if step

 

AGNOSTIC ALARMS LEDs will in-

6 is not done within 30

 

dicate normal operation. If the new

minutes).

 

calibration values, slope or con-

 

 

stant, is not within the parameters,

6 Push the CAL key; the CAL

 

the DIAGNOSTIC ALARMS LED

LED will be on solid. The

 

will indicate an alarm. (See Section

timer is activated for the sec-

 

7, TROUBLESHOOTING, for

ond calibration gas flow.

 

alarm codes). If the calibration was

When the timer times out, the

 

invalid, the Oxymitter 4000 will re-

CAL LED will flash a 2 pattern

 

turn to normal operation, as it was

flash or a 3 pattern flash (2

 

before a calibration was initiated,

pattern flash equals a valid

 

and the parameters will not be

calibration, 3 pattern flash

 

updated.

equals an invalid calibration).

 

(e) Place control loop in automatic.

If the slope or the constant is

 

 

 

out of specification, a diag-

6-3 LED STATUS INDICATORS

nostic alarm LED will be

 

6

flashing. The diagnostic alarm

a.

Diagnostic/Unit Alarms

will remain active until the

 

 

purge cycle is over. If the

Table 6-1 lists the types and status of

three pattern flash occurs

alarms that will be encountered. (See

without a diagnostic alarm,

Section 7, TROUBLESHOOTING, for a de-

the calibration gases could be

tailed description of each fault).

the same or the calibration

 

gas was not turned on.

b. When the electronics determines a calibra-

 

The CAL LED flashing indi-

tion is recommended, the CALIBRATION

RECOMMENDED LED is on solid.

cates the calibration is done.

 

(See Section 7, TROUBLE-

c. The CAL LED turns on when a calibration is

SHOOTING, for an explana-

recommended and is on during the calibra-

tion of the 2 pattern and 3

tion process. During calibration, the CAL

pattern flashes).

LED can be flashing, which would indicate

 

7 Remove the second calibra-

operator action is requested, or on solid,

which indicates calculations and measure-

tion gas and cap off the cali-

ments are in progress.

bration gas port.

 

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 6-5

Page 79
Image 79
Emerson Process Management IB106-340 LED Status Indicators, Nostic alarm LED will be, Flashing. The diagnostic alarm