Instruction Manual
Oxymitter 4000
b. Board Replacement | 10. Connect replacement board to board |
| (18 or 19). |
Perform the following procedure to replace power supply board (18, Figure
Disconnect and lock out power before working on any electrical components.
1.Turn off power to the system.
2.Remove screw (7) securing manifold cover lock (6) and remove the lock.
3.Remove manifold cover (14).
4.Remove two screws (11) attaching spacers (9) to manifold (5).
5.Being careful not to disconnect the board wiring, carefully lift power supply board (18) and interface board (19) from manifold (5) and set aside. Do not lose
6.Tag all leads on the board to be re- placed to simplify installation.
7.See Figure
See Figure
8.Remove stop nuts (22, Figure
9.Carefully separate boards (18 and 19).
11.Install screws (10), washers (21), and stop nuts (22) to secure power supply board (18) and interface board (19) to spacers (9).
12.Install all applicable leads in the appro- priate locations on the power supply board or interface board as shown in Figure
13.Install power supply board (18, Figure
14.Install manifold cover (14) and secure with manifold cover lock (6) and screw
(7).
c.Solenoid Replacement
The SPS 4000 manifold has a calibration gas 1 (high calibration gas) solenoid (20, Figure
Disconnect and lock out power before working on any electrical components.
1.Turn off power to the system.
2.Shut off the calibration gases at the cylinders.
3.Remove screw (7) securing manifold cover lock (6) and remove the lock.
4.Remove manifold cover (14).
5.Remove two screws (11) attaching spacers (9) to manifold (5).
6.Being careful not to disconnect the board wiring, carefully lift the board and spacer assembly from manifold (5) and set aside. Do not lose