m. Fault 13, Invalid Slope

The CALIBRATION LED flashes once,

pauses for three seconds, and repeats

(Figure 7-13).

1. During a calibration, the electronics

calculates a slope value. If the value of

TEST

POINTS

INC INC

HIGH LOW

GAS GAS

DEC DEC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS + PROCESS - % O2

TP1

J1

 

TP2

 

TP3

 

TP4

RED YEL GRN ORG

 

TP5

TP6

22220044

the slope is less than 35 mV/deg or

more than 52 mV/deg, the slope alarm

will be active until the end of the purge

cycle.

2. See paragraph 6-2,Calibration. Verify

the calibration by carefully repeating it.

Ensure the calibration gases match the

calibration gas parameters. If you at-

tach a multimeter to TP1+ and TP2-,

sample gas measurements are:

8% O2 23 mV

Figure 7-13. Fault 13, Invalid Slope

 

0.4% O2 85 mV

3.

Power down the Oxymitter 4000 and

 

remove it from the stack.

4.

Replace the cell per paragraph 6-8,

 

Cell Replacement.