Instruction Manual

IB-106-340 Rev. 2.4 April, 2001

Oxymitter 4000

 

0.125-27 NPT FEMALE

3.12 (79.25) MAX

 

 

 

OUTLET CONNECTION

 

 

NOTE: DIMENSIONS ARE IN INCHES WITH

 

1

 

 

2.250 (57.15)

 

MILLIMETERS IN PARENTHESES.

 

2

OUTLET

3

 

 

 

 

 

 

 

 

 

 

 

4.81

(122.17)

 

 

 

0.25-18 NPT FEMALE

 

 

 

 

 

 

 

 

 

 

 

 

INLET CONNECTION

8.50

 

 

 

 

 

 

 

 

(215.90)

FLOW SET

 

 

 

 

 

 

 

MAX

POINT KNOB

 

 

1.19

2.0

 

 

 

 

 

 

 

 

2 MOUNTING HOLES

 

 

 

(30.22)

 

 

 

 

(50.80)

 

 

 

 

3.19

(81.03)

LG

 

 

 

 

 

 

1.50THROUGH BODY FOR

 

 

DRAIN VALVE

 

 

 

 

 

 

 

 

(38.10)

0.312 (7.92) DIA BOLTS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

10.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(254)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.250 OR 6 MM O.D.

 

REF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TUBE COMPRESSION

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FITTING (SUPPLIED BY WECO)

1

FLOWMETER

 

0.2-2.0 SCFH

771B635H02

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

2" PRESSURE GAGE

0-15 PSIG

275431-006

 

 

 

TO ELECTRONICS

3

COMBINATION FILTER-REG.

0-30 PSIG

4505C21G01

 

 

 

 

 

 

 

 

 

OXYMITTER 4000

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SERIAL NO.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TAG NO.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

VOLTS: 85-264 VAC 48-62 Hz

WATTS: 500 VA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

OUTPUT: 4-20 mA LINE FUSE:

5 Amps

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.250 OR 6 MM O.D. TUBING (SUPPLIED BY CUSTOMER)

REF AIR SET

263C152G01

INSTRUMENT AIR SUPPLY

10-225 PSIG MAX PRESSURE

SCHEMATIC HOOKUP FOR REFERENCE AIR SUPPLY ON OXYMITTER 4000 PROBE HEAD.

26170035

Figure 2-11. Air Set, Plant Air Connection

2-4 PNEUMATIC INSTALLATION (FOR OXY- MITTER 4000 WITHOUT SPS 4000)

a.Reference Air Package

After the Oxymitter 4000 is installed, con- nect the reference air set to the Oxymitter 4000. The reference air set should be in- stalled in accordance with Figure 2-11.

Instrument Air (Reference Air): 10 psig (68.95 kPag) minimum, 225 psig (1551.38

kPag) maximum at 2 scfh (56.6 L/hr) maxi- mum; less than 40 parts-per-million total hydrocarbons. Regulator outlet pressure should be set at 5 psi (35 kPa). Reference air can be supplied by the reference air set of the IMPS 4000.

If using an IMPS 4000, refer to the IMPS 4000 Intelligent Multiprobe Test Gas Se- quencer Instruction Bulletin for the proper reference air connections.

2-14 Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 54
Image 54
Emerson Process Management IB106-340 Pneumatic Installation for OXY- Mitter 4000 Without SPS, Reference Air Package