l. Fault 12, EEPROM Corrupt

The O2 CELL LED flashes four times,

pauses for three seconds, and repeats

(Figure 7-12).

1. This alarm can occur if the EEPROM is

changed for a later version. At power

TEST

POINTS

INC INC

HIGH LOW

GAS GAS

DEC DEC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS +

PROCESS -

% O2

TP1 TP2 TP3 TP4

TP5 TP6

J1 RED YEL GRN ORG

22220043

 

up, the EEPROM is not updated.

2.

To correct this problem, power down

 

and then restore power. The alarm

 

should clear.

3.

If the alarm occurs while the unit is

 

running, there is a hardware problem

 

on the microprocessor board.

4.

If cycling the power does not clear the

 

alarm, see paragraph 6-5b,Electronic

 

Assembly Replacement.

Figure 7-12. Fault 12, EEPROM Corrupt

7

Rosemount Analytical Inc. A Division of Emerson Process Management

Troubleshooting 7-15

Page 113
Image 113
Emerson Process Management IB106-340 instruction manual Fault 12, Eeprom Corrupt