Figure 7-12. Fault 12, EEPROM Corrupt
l. Fault 12, EEPROM Corrupt
The O2 CELL LED flashes four times,
pauses for three seconds, and repeats
(Figure 7-12).
1. This alarm can occ ur if the EEPROM is
changed for a later version. At power
up, the EEPROM is not updated.
2. To correct this problem , power down
and then restore power. The alarm
should clear.
3. If the alarm occurs whil e the unit is
running, there is a hardware problem
on the microprocessor board.
4. If cycling the power does not c lear the
alarm, see paragraph 6-5b, Electronic
Assembly Replacement.