Oxymitter 4000

Instruction Manual

IB-106-340 Rev. 2.4

April, 2001

1

Signals:

 

Analog Output/HART

4-20 mA isolated from power supply, 950 ohms maxi-

 

mum load

Logic I/O

Two-terminal logic contact configurable as either an

 

alarm output or as a bi-directional calibration hand-

 

shake signal to IMPS 4000 or SPS 4000

 

Self-powered (+5 V), in series with 340 ohms

 

Conduit ports — 3/4 in.-14 NPT (one threaded hole for

 

both analog output and logic I/O)

Power Requirements:

 

Probe Heater

175 W nominal

Electronics

10 W nominal

Maximum

500 W

SPS 4000

 

Mounting

Integral to Oxymitter 4000

 

Remote from Oxymitter 4000

Materials of Construction:

 

Manifold/Electronics Enclosure

Aluminum

Mounting Brackets

316 stainless steel (SS)

Pneumatic Fittings

1/8 in. brass NPT (SS optional)

Pneumatic Tubing

1/4 in. Teflon (SS optional)

Assembly Hardware

Galvanized and stainless steel

Humidity Range

100% relative humidity

Ambient Temperature Range

-40°to 149°F (-40°to 65°C)

Electrical Classification

NEMA 4X (IP56)

Explosion-Proof Option (both pending)

CENELEC EExd IIB + H2

 

(Class 1, Div. 1, Group B,C,D)

Electrical Feedthroughs

1/2 in. NPT

Input Power

90 to 250 VAC, 50/60 Hz

Power Consumption

5 VA maximum

External Electrical Noise

EN 50 082-2, includes 4 KV electrostatic discharge

Handshake Signal to/from

 

Oxymitter 4000 (self-powered)

5 V (5 mA maximum)

Cal Initiate Contact Input from Control Room ....

5 VDC (self-powered)

Relay Outputs to Control Room

5 to 30 VDC, Form A (SPST)

 

(one “In-Cal”, one “Cal Failed”)

Cabling Distance between

 

SPS 4000 and Oxymitter 4000

Maximum 1000 ft (303 m)

Piping Distance between

 

SPS 4000 and Oxymitter 4000

Maximum 300 ft (91 m)

Approximate Shipping Weight

10 lbs (4.5 kg)

Fisher-Rosemount has satisfied all obligations coming from the European legislation to harmonize the product requirements in Europe.

Rosemount Analytical Inc. A Division of Emerson Process Management

Description and Specifications 1-11

Page 37
Image 37
Emerson Process Management IB106-340 Integral to Oxymitter, Remote from Oxymitter, 100% relative humidity, Nema 4X IP56