Instruction Manual

IB-106-340 Rev. 2.4 April, 2001

Oxymitter 4000

 

Table 5-1. Logic I/O Configuration

 

 

Mode

Configuration

 

 

0

The unit is not configured for any alarm condition.

 

 

1

The unit is configured for a Unit Alarm.

 

 

2

The unit is configured for Low O2.

3

The unit is configured for both a Unit Alarm and Low O2.

4

The unit is configured for a High AC Impedance/CALIBRATION RECOMMENDED.

 

 

5*

The unit is configured for both a Unit Alarm and a High AC Impedance/

 

CALIBRATION RECOMMENDED.

6

The unit is configured for both a Low O2 and High AC Impedance/CALIBRATION

 

RECOMMENDED.

7

The unit is configured for a Unit Alarm, a Low O2, and a High AC Impedance/

 

CALIBRATION RECOMMENDED.

8**

The unit is configured for a calibration handshake with IMPS 4000 or SPS 4000.

 

CALIBRATION RECOMMENDED will initiate the calibration cycle.

9

The unit is configured for a calibration handshake. CALIBRATION RECOM-

 

MENDED will not initiate the calibration cycle with the IMPS 4000 or SPS 4000.

*The default condition for an Oxymitter 4000 without an IMPS 4000 or SPS 4000. **The default condition for an Oxymitter 4000 with an IMPS 4000 or SPS 4000.

5-4 HART/AMS

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 68