Error Messages in Alphabetical Order

2-PORT CAL REQUIRED FOR AUX CHANNEL USE

.,. _

Emor Number l'hismessage is &splayed if you pressed ~~~~~~~~~without afu

217

2-port calibration being active. Perform (or recall) a full 2-port calibration and

 

/.,.,.. _;,.,.,...,.,...,.., .,..

_

.:...

 

set ~~~~~~~~~~Q~~, to &@ h the Ical] menu. ‘I&n you c-s enable a

 

..::...< :.;,.;~:;;.;;;;;;

~;~;.........;

.......-..:::::::.......:::::::;;>.,.,.:: ....<..<.;L.‘.

auxiliary channel.

ABORTING COPY OUTPUT

Information This message is displayed briefly if you have pressed m to abort a copy

Message operation. If the message is not subsequently replaced by error message number 25, PRINT ABORTED, the copy device may be hung. Press (Local) once more to exit the abort process and verify the status of the copy device. At this point, the copy device will probably have an error condition which must be 6xed. (For example: out of paper or paper jam.)

ADDITIONALSTANDARDSNEEDED

Error Number Error correction for the selected calibration class cannot be computed until you

6 8have measured all the necessary standards.

ADDRBSSKDTO TALK WITH NOTHING TO SAY

Error Number ybu have sent a read command to the analyzer (such as ENTER 716) without

31first requesting data with an appropriate output command (such as OUTPDATA). The analyzer has no data in the output queue to satisfy the request.

AIR FLOW RESTRICTED: CHECK FAN FILTER

Error Number Something is restricting the air flow into the analyzer. Check for any debris and

2 0clean or replace the fan filter.

ALL RRGISTBRS HAVE BEEN USED

Error Number You have used all of the available registers; you can store no more instrument

2 0 0 states even though you may still have sufficient memory. There are 31 registers available, plus the present instrument state.

1 O-2 Error Messages

Page 586