Basic Measurement Sequence and Example

Basic Measurement Sequence

There are five basic steps when you are making a measurement.

1. Connect the device under test and any required test equipment.

Caution Damage may result to the device under test if it is sensitive to the analyzer’s default output power level. To avoid damaging a sensitive DUT, be sure to set the output power appropriately before connecting the dut to the analyzer.

2.Choose the measurement parameters.

3.Perform and apply the appropriate error-correction.

4.Measure the device under test.

5.Output the measurement results.

Basic Measurement Example

This example procedure shows you how to measure the transmission response of a bandpass Glter.

Step 1. Connect the device under test and any required test equipment.

1. Make the connections as shown in Figure 2-l.

NETWORK ANALYZER

DEVICE UNDER TEST

pg6’e

Figure 2-1. Basic Measurement Setup

Step 2. Choose the measurement parameters.

2. Press w.

To set preset to “Factory Preset,n press:

Setting the Frequency Range.

3.To set the center frequency to 134 MHz, press: [centerl(TzJ(iiiJg

Meking Measurements 23