SWEEPTIMEINCREASED

Error Number You have made instrument changes that cause the analyzer sweep time to be

11automatically increased. Some parameter changes that cause an increase in sweep time are narrower IF bandwidth, an increase in the number of points, and a change in sweep type.

SWEEPTIMETOO FAST

Error Number The fractional-N and digital IF circuits have lost synchronization. Refer to the

12 HP 8753E N&work AnuZgzzr Semrice Guide for troubleshooting information.

SWEEPTRIGGERSETTO HOLD

 

 

Information

The instrument is in a hold state and is no longer sweeping. To take a new

Message

sweep, pres LMenu] ~~~~~~ ~~~~~~, or ~~~~~~~~~

 

;;;;;>:.;:

s...:.:::,....:...i~::::::..A..;;..;:.. ....AL. . ::...A./

i .-........:.....hw;;;.i; .,L.,.xA,7:::..::...... -

SYNTAXERROR

Error Number You have improperly formatted an HP-IB command. Refer to the HP87533

3 3Network Anal~/zer Programming and Command Reference oui& for proper

command syntax.

SYSTCTRL ORPASSCTRLINLOCALMENU

Error Number The analyzer is in talker/listener mode. In this mode, the analyzer cannot

3 6control a peripheral device on the bus Use the local menu to change to system controller or pass control mode.

TESTABORTED

Error Number You have prematurely stopped a service test. 113

THISLISTFREQ INVALID INHARM/3GHZRNG

Error Number Ybu have set frequencies in the list that are outside of the allowable frequency

133range for harmonic measurements, or are greater than 3 GHz on instruments without Option 006. Reduce the frequency range of the list.

Error Messeges 1O-26