TOO MANYNESTED SEQUENCES. SEQ ABORTED

Error Number You can only nest sequences to a maximum level of six. The sequence will

164 abort if you nest more than six.

TOO MANYSEGMENTS ORPOINTS

Error Number You can have a maximum of 30 segments or 1632 points in frequency list mode.

50In power meter calibrations, you can have a maximum of 12 segments for power sensor cal factors and power loss functions.

TRANSFORM, GATENOTALLOWED

Error Number You can perform a time domain transformation only in linear and CW sweep

16 types.

TROUBLE! CHECKSETUPAND STARTOVER

Service Error Your equipment setup for the adjustment procedure in progress is not correct. Number 115 Check the setup diagram and instructions HP 8753E Network Anulgm semrice

ouide. Start the procedure again.

WAITINGFORCLEAN SWEEP

Information In single sweep mode, the instrument ensures that aI.l changes to the

Message instrument state, if any, have been implemented before taking the sweep. The command that you have initiated is being processed and wiII not be complete until the new sweep is completed. An asterisk * is displayed in the left margin until a complete fresh sweep has been taken.

WAITINGFORDISK

Information This message is displayed between the start and finish of a read or write

Message operation to a disk.

lo-26 Error Messages