NO LIMITLINESDISPLAYED

Error Number You can turn limit lines on but they cannot be displayed on polar or Smith

144 chart display formats

NO MARKERDELTA -SPANNOTSET

Error Number You must turn the delta marker mode on, with at least two markers displayed,

15

::~:;~:~:~:.:~:.:.:

..... .:. ;.:..:.:.:.:.:.:.:.<....:~.:.:.:.:.,.:.;....

I; .::.,..:.:.~p,....

in order to use the ~~~~,~~~::: softkey function.

 

.:

i.:.: ..i. ~.........~.........~....... ..:

:..:

NO MEMORYAVAILABLEFORINTERPOLATION

Error Number You cannot perform interpolated error correction due to insufficient memory. 123

NO MEMORYAVAILABLEFORSEQUENCING

Error Number You cannot modify the sequence due to insufficient memory. 126

NO PHASELOCK: CHECKR INPUTLEVEL

Error Number The first IF’ signal was detected at pretune, but phase lock could not be

7acquired. Check the signal level to the R channel input to make sure it is

-35 dBm or higher. Refer to the HP 87533 N&work Anulgzer service Guide for troubleshooting.

NO SPACEFORNEW CAL. CLEARREGISTERS

Error Number You cannot store a calibration set due to insufficient memory. You can free

7 0more memory by clearing a saved instrument state from an internal register (which may also delete an associated calibration set, if all the instrument states using the calibration kit have been deleted). You can store the saved instrument state and calibration set to a disk before clearing them. After deleting the instrument states, press B to run the memory packer.

NOTALLOWEDDURINGPOWERMETERCAL

Error Number When the analyzer is performing a power meter calibration, the HP-B bus is

198 unavailable for other functions such as printing or plotting.

10.16 Error Messages

Page 600
Image 600