PROBEPOWERSHUT DOWN!

Error Number One or both of the probe power supplies have been shut down due to an

23 over-current, over-voltage, or under-voltage condition.

PROCESSINGDISPLAYLIST

Information The display information is being processed for a screen print to a copy device

Message and stored in the copy spool buffer. During this time, the analyzer’s resources are dedicated to this task (which takes less than a few seconds).

PWRMTRNOTON/CONNECTED ORWRONGADDRS

Error Number The power meter cannot be accessed by the analyzer. Verify that the power

117meter address and model number set in the analyzer match the address and model number of the actual power meter.

RANGECAUSEDPOWERLVLCHANGEINLIST

Error Number The selected power range changed the power level of one or more segments in

213 the swept list table. Change the segment power or change the power range.

REQUESTEDDATANOTCURRENTLYAVAILABLE

Error Number The analyzer does not currently contain the data you have requested. For

30example, this condition occurs when you request error term arrays and no calibration is active.

SAVEFAILED. INSUFFICIENTMEMORY

Error Number You cannot store an instrument state in an internal register due to insufficient

151memory. Increase the available memory by clearing one or more save/recall registers and pressing B, or by storing files to a disk.