INSUFFICIENTMEMORY, PWRMTRCAL OFF

Error Number There is not enough memory space for the power meter calibration array.

154Increase the available memory by clearing one or more save/recall registers, or by reducing the number of points.

INVALIDKEY

Error Number You pressed an undefined softkey. 2

LIMITTABLEEMPTY

Error Number Limit lines cannot be turned on unless a limit table has been created. Refer to

205“‘IWing a Device with Limit Lines” in Chapter 2 for information on how to create a limit table.

LISTMODE OFF: INVALIDWITHLO FREQ

Error Number List mode has been turned off in the frequency offset mode because it is

182 incompatible with your selected Lo frequency.

LISTTABLEEMPTY

Error Number The frequency list is empty. To implement list frequency mode, add segments

9 to the list table.

LOGSWEEPREQUIRES2 OCTAVEMINIMUMSPAN

Error Number A logarithmic sweep is only valid if the stop frequency is greater than four

150times the start frequency. For frequency spans of less than two octaves, the sweep type automatically reverts to linear sweep.

LOWPASS: FREQLIMITSCHANGED

Information The frequency domain data points must be harmonically related from dc to the

Message stop frequency. That is, stop = n x start, where n = number of points If this condition is not true when a low pass mode (step or impulse) is selected and transform is turned on, the analyzer resets the start and stop frequencies. The stop frequency is set close to the entered stop frequency, and the start frequency is set equal to stop/n.

1O-14 Error Messages