D2/DlINVALIDWITHSINGLECHANNEL

Error Number You can only make a D2Dl measurement if both channels are on. 130

D2/DlINVALID: CHlCH2NWIPTSDIFFERENT

Error Number You can only make a D2/Dl measurement if both channels have the same

152 number of points.

DEADLOCK

Error Number A fatal firmware error occurred before instrument preset completed. Call your

111 local Hewlett-Packard sales and service office.

DEMODULATIONNOTVALID

Error Number Demodulation was selected when the analyzer was not in CW time mode.

17 Select demodulation only after putting the analyzer into CW time mode.

DEVICE:noton,notconnect, wrongaddrs

Error Number The device at the selected address cannot be accessed by the analyzer. Verify

119that the device is switched on, and check the HP43 connection between the analyzer and the device. Ensure that the device address recognized by the analyzer matches the HP-IB address set on the device itself.

DIRECTORYFULL

Error Number There is no room left in the directory to add flies. Either delete files or get a

188 new disk.

DISKHARDWAREPROBLEM

Error Number The disk drive is not responding correctly. Refer to the HP 8753E Network

39Andper semrice Guide for troubleshooting information. If using an external disk drive, refer to the disk drive operating manual.

1 O-8 Error Messafles

Page 592