ATN (Attention)

The active controller uses this line to define whether

 

the information on the data bus is command-oriented or

 

data-oriented. When this line is true (low), the bus is in the

 

command mode, and the data lines carry bus commands. When

 

this line is false (high), the bus is in the data mode, and the

 

data lines carry device-dependent instructions or data.

SRQ (Service Request)

This line is set true (low) when a device requests service

 

and the active controller services the requesting device. The

 

network analyzer can be enabled to pull the SRQ line for a

 

variety of reasons such as requesting control of the interface,

 

for the purposes of printing, plotting, or accessing a disk.

REN (Remote Enable)

This line is used exclusively by the system controller. When this

 

line is set true (low), the bus is in the remote mode, and devices

 

are addressed by the controller to either listen or talk. When

 

the bus is in remote mode and a device is addressed, it receives

 

instructions from the system controller via HP-IB rather than

 

from its front panel (pressing ILocal) returns the device to

 

front-panel operation). When this line is set false (high), the

 

bus and ail of the connected devices return to local operation.

EOI (End or Identify)

This line is used by a talker to indicate the last data byte in a

 

multiple-byte transmission, or by an active controller to initiate

 

a parallel-poll sequence. The analyzer recognizes the EOI line

 

as a terminator, and it pulls the EOI line with the last byte of a

 

message output (data, markers, plots, prin% error messages).

 

The analyzer does not respond to parallel poll.

HP-IB Requirements

Number of Interconnected Devices:15 maximum.

Interconnection Path Maxims Cable Length: 20 meters maximum or 2 meters per device

 

(whichever is less).

Message Transfer Scheme:

Byte serial, bit parallel a synchronous data

 

transfer using a 3-line handshake system.

Data Rate:

Address Capability:

Multiple-Controller Capability:

Maximum of 1 megabyte-per-second over the specified distances with t&state drivers. Actual data rate depends on the transfer rate of the slowest device connected to the bus

Primary addresses: 31 talk, 31 listen. A maximum of 1 talker and 14 listeners can be connected to the interface at given time.

In systems with more than one controller (like this instrument), only one controller can be active at any given time. The active controller can pass control to another controller, but only the system controller can assume unconditional control. Only one system controller is allowed.

11-16 Compatible Peripherals