HP 8753E manual Verify performance, Cal kit menu. It will be saved with calibration sets

Models: 8753E

1 699
Download 699 pages 49.83 Kb
Page 372
Image 372

Label Class Menu

The label class menus are used to dellne meaningful labels for the calibration classes. These then become softkey labels during a measurement calibration. Labels can be up to ten characters long.

Label

Kit Menu

 

. . . . ..

;.

This ::U#l#L%JtT softkey within the modify cal kit menu, provides access to this menu. It

.:.i ~.~.~..~..i

i //..

is identical to the label class menu and the label standard menu described above. It allows deilnition of a label up to eight characters long.

After a new calibration kit has been dellned, be sure to specify a label for it. Choose a label

that describes the connector type of the calibration devices. ‘Ihis label will then appear in the

.;: ‘y= <s;<.:<;;;...;p:

:A.

..;;;;;-“;,,, ,i, :

label h the come&ion menu ad the ~~~~~~~~:~~g.i label h the se&&

~~~~~~/fE,.~1’;~,;, softkey

:<..:.,,. i..,.....A...v;:....i..~

~..i

_:.../.i;;;;;;

/..::.. ..:................i s.2;;:

.s.......>.....ss..ii.......ASi..

cal kit menu. It will be saved with calibration sets.

Verify performance

Once a measurement calibration has been generated with a user-dehned calibration kit, its performance should be checked before making device measurements. lb check the accuracy that can be obtained using the new calibration kit, a device with a well-deiined frequency response (preferably unlike any of the standards used) should be measured. The verification device must not be one of the calibration standards: measurement of one of these standards is merely a measure of repeatability.

lb achieve more complete verification of a particular measurement calibration, accurately known verification standards with a diverse magnitude and phase response should be used. National standard traceable or HP standards are recommended to achieve verifiable measurement accuracy.

Note

The published specifications for the HP 8753E network analyzer system

 

include accuracy enhancement with compatible calibration kits Measurement

 

calibrations made with user-defined or modified calibration kits are not subject

 

to the HP 8753E specifications, although a procedure similar to the system

 

verification procedure may be used.

 

 

Application and Operation Concepts 6-61

Page 372
Image 372
HP 8753E Verify performance, Cal kit menu. It will be saved with calibration sets, Verification procedure may be used