NETWORli A N A L Y Z E R

pg624e

Figure 3-3. B Channel External Connection

4. Measure the output power in the R channel by pressing:

Observe the 13 to 16 dR offset in measured power. The actual input power level to the R channel input must be 0 dBm or less, -10 dRm typical, to avoid receiver saturation effects The minimum signal level must be greater than -35 dBm to provide sufficient signal for operation of the phaselock loop.

5.You cannot trust R channel power settings without knowing about the offset involved. Perform a receiver calibration to remove any power offsets by pressing:

Once completed, the R channel should display 0 dRm. Changing power ranges will require a recalibration of the R channel.

Mating Mixer Measurements 3-S